WATN2001 TOTAL NITROGEN ANALYZER

WATN2001 TOTAL NITROGEN ANALYZER

Features

  • Core components are imported, high reliability, long lifetime, suitable for long time online monitoring
  • All part contact with liquid will be particularly tested to ensure a long service life
  • Specially designed flow path to prevent blocking
  • Photoelectric measurement method, accurate sampling, low sample injection error
  • Automatic chromaticity and turbidity compensation algorithm give fill consideration to the situation of actual water samples, real and reliable monitoring results.
  • Unique hybrid technology or bubbles agitation ensures thorough mixing of samples and reagents
  • When a reagent leaks automatic there will be an alarm and reminders for maintenance.
  • Equipment has self-diagnosis function, if there is any defects and alarm will be alerted and display on screen.

Application

  • Sewage treatment plants
  • Industrial production of water
  • Wastewater treatment process
  • Drinking water industry
  • Process monitoring in industrial facilities
Contact Us

 

Catalogue

Description

WATN 2001 is an accurate and reliable total nitrogen analyzer.
Under temperatures of 60°C and above, potassium persulfate is completely decomposed into potassium bisulfate and oxygen. This followed by the reaction at 120°C which causes the nitrogen compounds and all organic matter in the water sample to be oxidized or decompose. This final state allows the absorbance of water samples to be measured by UV spectrophotometry at the a fixed range. Allowing the total nitrogen content to be calculated according to the absorbance